This equipment is 8 channels per set, and 10 sets for one rack, 80 channels.
Battery type can be tester: pack tester for Li-ion, LiFePO4, Ni-MH, Ni-CD and lead acid battery.
Items can be tested:
Current, voltage, power, DCIR,capacity, cycle test, Pulse test, etc
Reliable hardware with independent constant current and voltage control.
8 Channels Battery Tester Hardware Features
1)According to the international standard design, each channel of the system has an independent constant-current source and a hardware constant-voltage source which make system stable & reliable
2)Modular design leads to easy maintain
3)Different holder selectable, satisfy your different needs
4)Efficient heat dissipation system, ensure proper working of equipments under 24hrs full load condition
5)Double loop-control adopted,no peak occured from CC to CV, protect the battery
8 Channels Battery Tester Software Features
1)Each channel available for indepent program testing,Step continuable after Pause in any conditions(including blackout)
2)Host computer adaptable to all kinds of lower machine,with interactive interface,easy to operate
3)Fast data acquisition,sensitive to the changing of current & voltage,(AD ,16bit),dairy function available
4)Powerful for data and chart processing ,support four types of report with various form formats
Input Power AC 220V ±10% / 50Hz
Input Active Power 365 W
Resolution AD:16bit;DA:16bit
Input Resistance ≥1MΩ
Voltage Per Channel Constant Voltage Control Range 0.05V~10V
Lowest Discharge Voltage 2.5V
Accurancy ± 0.1% of FS
Stability ± 0.1% of FS
Current Per Channel Current Range 0.015A~3A
Accurancy ± 0.1% of FS
Constant Voltage Stop Current 0.006A
Stability ± 0.1% of FS
Power Per Channel Output Power 30W
Stability ± 0.2% of FS
Time Current Response Time Maxmum Current ≤20ms
Data Recording Record Condition Minimum Time Interval:100ms
Minimum Voltage Interval:20mV
Minimum Current Interval:6mA
Record Frequency 10Hz
Charge Charge Mode Constant Current Charge、Constant Voltage Charge、Constant Curren and Voltage Charge、Constant Power Charge
End Condition Voltage、Current、Relative Time、Capacity
Discharge Discharge Mode CCD、CPD、CRD
End Condition Voltage、Current、Relative Time、Capacity、Voltage Interval
Pulse Test Charge CCC CCP
Discharge CCC CCP
Minimum Pulse Width 500ms
Pulse number 32 diffirent pulse for a single step
Charge and Discharge Constant Switch Support continuous switched in a single pulse step
End Condition Voltage、Relative Time
DCIR Test Support
Cycle Loop Measure Range 1~65535 times
Max Steps Per Loop 254
Nested Loop Nested loop Function,Max Support 3 Layers
Protection Software Protection ● Power-down Data Protection
● Off-line Testing Function
● Setting Protection Condition,Setting Parameter:Low Voltage limit、Upper Voltage limit、Low Current limit、Upper Current limit、Delay Time
Hardware Protection Additional Anti-reverse Protection Model
Channel Control Independent control per channel
Voltage and Current Testing Sample 4-wire connecting
Noise <85dB
Database MySQL
Upper Machine Cummunication mode Based on TCP/IP
Data Output EXCEL2003,2010、TXT
Server disk configuration 500GB
Leak Current 0.01mA
Server Operating System Windows XP,Windows 7
Communication Interface Internet Access
****Different Specifications could be customized ,Ranging from 1mA-3000A, 5-800V;)****
If you are interested in NEWARE battery testing equipment, pls contact us !
Elley Wang Mobile:15913168930 Skype:elleywangyao Email: elley.wangyao@gmail.com / wangyao@neware.com.cn